Introduction

The CEL-240 is a completely new design of sound level meter featuring the power of Digital Signal Processing (DSP) technology and the simplicity and low cost of a traditional analog instrument. It can be used for a wide range of noise measurement tasks and is suitable for many everyday tasks. Complying with all the ANSI and IEC accuracy specifications in the Type 2 category the CEL-240 will appeal to new and experienced users alike.

Key Benefits

  • Simple 2 button operation

  • Wide range from 30 to 130dB

  • Large backlit graphic LCD screen 128 x 128 pixels

  • Slow, Fast and Impulse time weightings

  • Current and highest noise level shown in display

  • Large numeric and analog bar graph display

  • Output to computer for simple data logging with PC

  • Auto calibration at selected decibel level

  • Scrolling sound level display

Applications

The CEL-240 is an ideal instrument for industrial hygiene measurements of noise in the workplace. The meter has a low and high measurement range that covers a full 70dB on each range. It can be used for machinery noise surveys as well as all general purpose workplace noise level measurements. With the A and C frequency scales and the Slow time response the meter can be used to assess the right hearing protectors using the NRR method. A non-decaying maximum hold feature on every display captures and displays the highest sound pressure level of any noise until reset by the user.

Features

The CEL-240 features an automatic back light for the display that comes on for 30 seconds whenever a key is pressed. A standard ¼” threaded socket is provided on the back of the meter to allow it to be mounted on a tripod for fixed applications. The meter is powered from 3 x AA alkaline batteries and will run for up to 35 hours or continuously when powered from mains to USB style adaptor. Digital and analog outputs are available via a USB mini A or 2.5mm jack sockets to connect the CEL-240 to other external computers or recorders or to the dB24 software package.

Technical Specification - General

Accuracy: ANSI S1.4 Type 2, IEC 61672-1 2002-5
Microphone type: ¼” Electret mic. in standard ½” fixed housing ¼” Electret mic. in standard ½” fixed housing
Reference Conditions: 68°F (20°C) air temperature,
65% Relative Humidity,
1013 mbar (101.325 kPa) atmospheric pressure.
Operating Temperature Range: 32 to 104°F (0 to 40°C) (Class 2)
Effect of Humidity: Less than ±0.5dB over the range 30 to 90% RH (non-condensing), rel. to value at ref. conditions
Operating pressure range: 650 to 1080 mbar (65 to 108 kPa)
Batteries: 3 x AAA Alkaline or rechargeable types
Battery Life: (hours) 20 hours without backlight
Dimensions w x h x d: (in/mm) 2.8 x 9.0 x 1.2 in (71.5x 230.0x 31.0mm) including preamplifier and microphone
Controls: 3 buttons for power and context sensitive menus selection with initial regional configuration screen
 

Technical Specification - Performance

Total measurement range (dB) 30 to 130
Dynamic span on single range (dB) 70
Number of measurement ranges - 2 ranges (Low and High dB) (30 - 100 & 60 - 130)
Noise floor (A weighted dB) < 33
Frequency weightings A and C
Time weightings Slow, Fast and Impulse
Displayed parameters Instantaneous level - Lp, Maximum level - Lmx
Display type 128 x 128 dot matrix LCD digital including real-time analog bar graph scale
Display resolution - numeric (dB) 0.1
Display resolution - graphical (dB) 1
Update rate for display (seconds) 0.5
Displayed time span for time history chart (minutes) Last 1 or 5
Backlight for display 30 sec after key press
Calibration method Automatically recognized by meter
Signal detected when calibrator placed over microphone Display level set to 114.0 dB
External power option (5 Vdc) Yes via CEL- XXXX PSU via USB socket
Analog outputs AC (and optional DC) via 2.5 mm jack socket
AC output characteristics - (Provided for DAT tape / PC wave file recording or headphone applications) Approx 0.5V RMS FSD output on selected sound level measurement range. Minimum load impedance 22k .
DC output characteristics - (Provided at time of order as option for connection to chart recorder or pc data logging system 0 to 3.3V DC for FSD on selected range. Output corresponds to selected frequency and time weighting. 2k Output impedance
Digital output USB 2.0 format of instantaneous sound level via 'mini A' USB socket (also powers meter)
Digital output characteristics - (value output once per second) Instantaneous SPL output (software required) as per selected frequency and time weightings.

Ordering information

CEL-240 Digital Sound Level Meter Type 2 with standard accessories
CEL-110/2 Acoustic calibrator Class 2 including ½" to ¼" microphone coupler
CEL-4627 Foam windscreen to protect against wind induced interference
CEL-240/K1 Standard sound level meter kit including meter, calibrator, windscreen and kit case plus various accessories, batteries, etc
CEL-240/K2 All items in K1 kit plus CMC51 USB cable to pc plus dB24 software
CEL-6840 Attaché kit case for meter and standard accessories
CEL-6842 dB24 software for data logging in spreadsheet format file
CMC51 Universal power supply at 5V via USB mini A cable from computer
Back

Return to SMG/interlink®
For more information  Email Mail  SMG/Interlink®
Interlink Marketing Corporation
dba SMG/interlink®
PO Box 260795
Plano, Texas 75026
Toll free 1-800-804-6196
Ph: 972-985-0883
Fx: 972-985-3252
2203 Glen Forest Ln
Plano, Texas 75023

credit card logos

Copyright © 1997-2012 Interlink Marketing Corporation

SiteLock