Definer™ 1020 Primary Gas Flow Calibrator
5 slm to 500 slm
(±0.25% of reading volumetric at STP)
The
Definer™ 1020 is a compact
and powerful calibrator featuring Bios' patented and proven DryCal® technology.
The Definer 1020 calibrator
has
an operating range from 5 to 500 liters per minute. Up to ten
times the volume of any other piston prover on the market! at an
accuarcy of ±0.25% of reading!
The portability of this new model, which weighs in at a "feathery"
90lbs. is a dramatic and welcome improvement over the cumbersome
bell provers, used in the past.
Flow Ranges
- 5-500 sL/min
Accuracy
- 5-500 sL/min ±0.25%
of reading*
Weight -
90 lbs/ 41kg
Dimensions
(H x W x D) - 24W" x 30H" x 12D"
Flow measurements -
Single, Continuous or Burst modes
Flow
measurements in the average, 1 - 100 measurements
* At a gas pressure of
760 mmHg, and a gas temperature of 25° centigrade
with standardization temperature set to 21.1° centigrade.
User-Selectable
Measurement Units
Volumetric Flow
L/min, cf/min
Standard Flow
sL/min, scf/min
Pressure
mm Hg, PSI , kPa
Temperature
°C, °F
Bios
Definer 1020 Specifications
Data Cable:
1 meter,
for use with data port, Bios Optimizer™ software,
and PC serial port
Data Port:
RS-232 (serial),
for use with data cable and Bios Optimizer™ software
Display:
Graphical LCD with backlight
Flow Measurement Cell:
Integrated (all-in-one design; no additional flow models)
Flow Measurement Style:
Single (manual), Continuous or Burst, with averaging function
user-selectable from 1 to 100
Flow Measurement Type:
Volumetric (primary by nature), Standardized
Flow Mode:
Pressure or suction
Gas Compatibility:
Noncorrosive, noncondensing, noncombustible gases, less than
70% humidity
Humidity (Ambient):
0-70%, noncondensing
Inlet and Outlet Fittings:
1 ½” Swagelok® compression
fitting
Overpressure Release Fitting:
½” Swagelok® compression
fitting
Storage Temperature:
0–70° C
Power:
External power module
Input: 100-240 VAC, 1.6A (max), 50-60 Hz
Output: 12 VDC, 3.0 A
Pressure (Operating):
10 - 19.5 PSI (absolute)
Temperature (Operating):
15-35°C
Temperature Sensor:
Accuracy ±0.03%
(max), measured in the gas flow stream
Return